Laboratory Airflow Monitors & Controls
TYPE C STANDARD &
NIGHT SETBACK
LED & DIGITAL - VAV SYSTEMS
Specifications and Owner’s Manual
2216 60th Drive East • Bradenton, Florida 34203
(941) 751-7596 • Fax (941) 751-7598 • www.plastecvent.net
PLASTEC
®
VENTILATION, INC.
Vista de pagina 0
1 2 3 4 5 6 ... 13 14

Indice de contenidos

Pagina 1 - NIGHT SETBACK

Laboratory Airflow Monitors & ControlsTYPE C STANDARD &NIGHT SETBACKLED & DIGITAL - VAV SYSTEMSSpecifications and Owner’s Manual2216 60t

Pagina 2 - General Description

10Overall View2216 60th Drive East • Bradenton, Florida 34203(941) 751-7596 • Fax (941) 751-7598 • www.plastecvent.netFront plate/fasciaChemical

Pagina 3 - Operating Principle

Alarm Relay220v 3A(optional)ONOFFSettings:1. On: Alarm always active Off: Alarm disabled when fan is off2. Alarm delay setting: On: 30 seconds Off:

Pagina 4 - TYPE C STANDARD

INSTALLATION1) Drill a Ø 23mm / 0.91" hole either in the top or side of the fume cupboard to allow sensor positioning. Make sure sensor is not in

Pagina 5

VELOCITY TYPETo change Digital readout from/to m/s to FPM:Press simultaneously the following buttons: until buzzer sounds (about 5 seconds).DISP

Pagina 6

MAINTENANCEFront plate and stickers of airflow monitor may be cleaned with mild soap and water on a damp cloth to remove finger marks, oils and residu

Pagina 7

General DescriptionBENEFITS• Safety: air flow is maintained at required speed at all sash heights• Energy savings: only the lowest required amount of

Pagina 8 - Specifications

When the fume extraction fan is running, it causes negative pressure inside the fume cupboard. If the sash is lowered, the negative pressure become mo

Pagina 9

115 VAC /12 VDCPower SupplyControllerSensorSensorSignal InputSash High Alarm InputSash movement=Pressure decrease or increaseHood Lighton/off relayAir

Pagina 10 - TYPE C NIGHT SETBACK

115 VAC /12 VDCPower SupplyControllerSensorSensorSignal InputSash High Alarm InputSash movement=Pressure decrease or increaseHood Lighton/off contactN

Pagina 11 - Contacts & Features

5AirControl C LEDPart Number 819703 819704 1 green LED for normal air speed 1 green LED for normal air speedDisplay – Visual 1 red LED flashing f

Pagina 12 - CONTROLLER SET UP PROCEDURE

AirControl C LEDPart Number 819703 819704 1 green LED for normal air speed 1 green LED for normal air speedDisplay – Visual 1 red LED flashing fo

Pagina 13 - FACTORY DATA RESET

MONITOR & MOUNTING BOXMounting Box90 mm / 3.54"210 mm / 8.27"SENSOR65 mm / 2.56"28 mm / 1.10"40 mm / 1.57"Ø 22 mm / 0.87&

Pagina 14 - WARRANTY

7Chemical and pressureresistant stickerFlashing red LED foralarm: air speed too lowAlarm Mute push buttonCustomizableResin StickerDigital display.Not